​The energy loss through the sample cell is compared with the reference signal provided by the filter wheel to produce a signal proportional to concentration, with little effect from interfering gases within the sample. This design produces excellent zero and span stability and high signal to noise ratio, allowing excellent performance over a wide concentration range.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.

​The Model T360 comes with NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software​.​​​​​

​​The following table provides a list of items that need regular maintenance to ensure your product remains in good working condition.

Maintenance

ITEM ACTION​
FREQUENCY CAL CHECK REQUIRED
Particulate FilterReplaceWeekly or as neededYes
Verify Test FunctionsRecord and analyzeWeekly or after any Maintenance or Repair No
Pump DiaphragmReplaceAnnuallyYes
Perform Flow CheckCheck FlowAnnually No
Perform Leak CheckVerify Leak TightAnnually or after any Maintenance or RepairNo
Pneumatic linesExamine and cleanAs neededYes if cleaned
CleaningCleanAs neededOnly if cover removed
​​​​​​