​Paramagnetic technology is non-depleting, which means that there are no consumable parts, and ensures a consistent performance over time. The selectivity of the paramagnetic measurement for oxygen means that there is little or no interference from other common background gases. The T802 offers a stable and inherently linear measurement of oxygen, making it possible to calibrate the analyzer by checking only two points. There is no requirement for a reference gas during operation.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.

The Model T802 comes with the NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.​​

​​The following table provides a list of items that need regular maintenance to ensure your product remains in good working condition.

Maintenance

ITEM ACTION FREQUENCY CAL CHECK MANUAL SECTION
Particulate Filter Replace Weekly or as needed No 10.3.1
Verify Test Functions Record and analyze Weekly or after any Maintenance or Repair No 11.1.2
Pump Diaphragm Replace Annually Yes 10.3.2
Perform Flow Check Check Flow Annually No 10.3.4
Perform Leak Check Verify Leak Tight Annually or after any Maintenance or Repair No 10.3.3
Pneumatic lines Examine and clean As needed Yes if cleaned
Chassis Wipe down As needed Only if cover removed
​​​​​​​​