Model T102 with NumaView™ Software​

UV Fluorescence Total Reduced Sulfur Analyzer​​​

​The Model T101 H2S analyzer uses the proven UV fluorescence principle to measure hydrogen sulfide at levels commonly required for ambient air monitoring.

The Model T101 is equipped with an internally mounted catalytic converter set at 315°C to convert H2S to SO2. A switching mode alternately measures H2S and SO2 while showing both readings concurrently on the front display.

Model T102 TRS Analyzer Front View

​​​


All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. All instrument set up, control and access to stored data and diagnostic information is available through the front panel using NumaView™ Software, or via RS232, Ethernet, or USB com ports, either locally or by remote connection using the included NumaView™ Remote PC Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.

​​The Model T102​ comes with NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.​​​​​​​​​​​

Features & Applications

​​The following table provides a list of items that need regular maintenance to ensure your product remains in good working condition.

Maintenance

ITEM ACTION FREQUENCY CAL CHECK MANUAL SECTION
SO2 scrubberReplace As required Yes7.3.3
H2S à SO2 Converter CatalystReplace As required Yes7.3.5
1Particulate filterChange particle filterWeeklyNo7.3.1
Verify test functionsReview and evaluateWeeklyNoAppendix C in manual
Zero/span checkEvaluate offset and slopeWeekly -- 6.3, 6.6, 6.9
1Zero/span calibrationZero and span calibrationEvery 3 months -- 6.2, 6.4, 6.5, 6.7, 6.8
1External zero air scrubber (optional)Exchange chemicalEvery 3 monthsNo7.3.4
1Perform flow checkCheck FlowEvery 6 MonthsNo8.5.2
1Critical flow orifice & sintered filtersReplaceAnnuallyYes7.3.7
Internal IZS Permeation TubeReplaceAs requiredYes7.3.2
Perform pneumatic leak checkVerify Leak TightAnnually or after repairs involving pneumaticsYes8.5.1
2Pump diaphragmReplaceAt least Every 2 years or if pres is ≥ 33.00 in-Hg-AYes​See instruction in diaphragm kit 
PMT sensor hardware calibrationLow-level hardware calibrationOn PMT/ preamp changes  if  0.7 < SLOPE or  SLOPE >1.3Yes8.6.4
1 These items are required to maintain full warranty; all other items are strongly recommended.​
2 A pump rebuild kit is available from Teledyne API's Technical Support, including all instructions and required parts.​