Exceptional stability is achieved with the use of an optical shutter to compensate for PMT drift and a reference detector to correct for changes in UV lamp intensity. A hydrocarbon ‘kicker’ and advanced optical design combine to prevent inaccuracies due to interferents.

​​All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. All instrument set up, control and access to stored data and diagnostic information is available through the front panel using NumaView™ Software, or via RS232, Ethernet, or USB com ports, either locally or by remote connection using the included NumaView™ Remote PC Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.

​​The Model T100 comes with NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.​​​​​​​​​​​

​​The following table provides a list of items that need regular maintenance to ensure your product remains in good working condition.

Maintenance

ITEM ACTION FREQUENCY CAL CHECK MANUAL SECTION
1Particulate filter Change particle filter Weekly No 10.3.1
Verify test functions Review and evaluate Weekly No 10.2; Appendix C in manual
Zero/span check Evaluate offset and slope Weekly --  9.3, 9.6, 9.9
1Zero/span calibration Zero and span calibration Every 3 months --  9.2, 9.4, 9.5, 9.7, 9.8
1External zero air scrubber (optional) Exchange chemical Every 3 months No 10.3.3
1Perform flow check Check Flow Every 6 Months No 10.3.7
Internal IZS Permeation Tube Replace Annually Yes 10.3.2
Perform pneumatic leak check Verify Leak Tight Annually or after repairs involving pneumatics Yes 10.3.6
2Pump diaphragm Replace Annually Yes Refer to diaphragm kit instructions
Calibrate UV Lamp Output Perform lAMP​ CAL  Prior to zero/span calibration or PMT hardware calibration --  5.9.6 & 11.7.2.5
3PMT sensor hardware calibration Low-level hardware calibration On PMT/ preamp changes  if 0.7
< slope or slope >1.3
Yes 11.7.2.8
1Sample chamber optics Clean chamber, windows and filters As necessary Yes 11.7.2.2 & 11.7.2.3
1Critical flow orifice & sintered filters Replace As necessary Yes 10.3.4​
1 These items are required to maintain full warranty; all other items are strongly recommended.​
2 A pump rebuild kit is available from Teledyne API's Technical Support, including all instructions and required parts.​
3 Replace desiccant bags each time the inspection plate for the sensor assembly is removed.​​​